Walk-up NMR user instructions for Bruker: Difference between revisions
Jump to navigation
Jump to search
| Line 3: | Line 3: | ||
==Step by step== | ==Step by step== | ||
# log in to workstation | # log in to workstation. | ||
# log in to TopSpin | # log in to TopSpin. | ||
# type ej on the command line or click the eject button to eject the current sample that is in the magnet | # type ej on the command line or click the eject button to eject the current sample that is in the magnet. | ||
# remove the old sample from the top of the magnet | # remove the old sample from the top of the magnet. | ||
# remove the old sample from the sample holder | # remove the old sample from the sample holder. | ||
# place your sample in the sample holder | # place your sample in the sample holder. | ||
# slide your sample down in the holder until it hits the stop | # slide your sample down in the holder until it hits the stop. | ||
# place the sample/holder on top of the column of air on the magnet | # place the sample/holder on top of the column of air on the magnet. | ||
# go back to the workstation, click insert sample and wait for the sample to drop down into the probe | # go back to the workstation, click insert sample and wait for the sample to drop down into the probe. | ||
# type lock on the command line | # type lock on the command line. | ||
# when the list of solvents appears, choose the one that your sample is in. Wait for locking. | # when the list of solvents appears, choose the one that your sample is in. Wait for locking. | ||
# once the sample is locked, type topshim on the command line to shim the sample | # once the sample is locked, type topshim on the command line to shim the sample. | ||
# while the sample is being automatically shimmed, type edc on the command line to create a new dataset | # while the sample is being automatically shimmed, type edc on the command line to create a new dataset. | ||
# edc opens the window that sets everything for the experiment to run your sample | # edc opens the window that sets everything for the experiment to run your sample (example [[New Dataset Window for Bruker TopSpin|here]]). | ||
Revision as of 10:26, 13 May 2020
Introduction
If you need to use a Bruker NMR instrument for a liquid sample, here are some quick instructions to get your sample running as easily as possible:
Step by step
- log in to workstation.
- log in to TopSpin.
- type ej on the command line or click the eject button to eject the current sample that is in the magnet.
- remove the old sample from the top of the magnet.
- remove the old sample from the sample holder.
- place your sample in the sample holder.
- slide your sample down in the holder until it hits the stop.
- place the sample/holder on top of the column of air on the magnet.
- go back to the workstation, click insert sample and wait for the sample to drop down into the probe.
- type lock on the command line.
- when the list of solvents appears, choose the one that your sample is in. Wait for locking.
- once the sample is locked, type topshim on the command line to shim the sample.
- while the sample is being automatically shimmed, type edc on the command line to create a new dataset.
- edc opens the window that sets everything for the experiment to run your sample (example here).