Walk-up NMR user instructions for ThermoFisher: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
No edit summary |
||
| (6 intermediate revisions by the same user not shown) | |||
| Line 1: | Line 1: | ||
The picospin tabletop NMR instruments utilize an injection system for samples rather than the usual 5mm NMR tubes. The instrument also has a web-based interface for control and processing. | The [[NMR Instrument Manufacturers|picospin]] tabletop NMR instruments utilize an injection system for samples rather than the usual 5mm NMR tubes. The instrument also has a web-based interface for control and processing. Here are some instructions to get a quick spectrum of your sample: | ||
# Fill a 1mL plastic syringe approximately half way. Be sure to remove any air bubbles that may be present. | |||
# Check to make sure a filter is properly connected and a collection flask is at the "out" port. | |||
# Slowly inject your sample into the "in" port until you observe a few drops coming out of the "out" port. | |||
# Go to the picospin page of the web browser interface | |||
# Select the '''Files''' option. | |||
# Select the '''Standard Run''' option. | |||
# Click '''Use these settings at run'''. | |||
# The '''Run''' page will appear. | |||
# Check the transmitter frequency and set the scans. | |||
# Click '''Start Run'''. | |||
# Process spectra with MNova, which is included with the product. | |||
[[File:picospin.jpg|400px]] | |||
Latest revision as of 03:37, 25 May 2020
The picospin tabletop NMR instruments utilize an injection system for samples rather than the usual 5mm NMR tubes. The instrument also has a web-based interface for control and processing. Here are some instructions to get a quick spectrum of your sample:
- Fill a 1mL plastic syringe approximately half way. Be sure to remove any air bubbles that may be present.
- Check to make sure a filter is properly connected and a collection flask is at the "out" port.
- Slowly inject your sample into the "in" port until you observe a few drops coming out of the "out" port.
- Go to the picospin page of the web browser interface
- Select the Files option.
- Select the Standard Run option.
- Click Use these settings at run.
- The Run page will appear.
- Check the transmitter frequency and set the scans.
- Click Start Run.
- Process spectra with MNova, which is included with the product.