The picospin tabletop NMR instruments utilize an injection system for samples rather than the usual 5mm NMR tubes. The instrument also has a web-based interface for control and processing. Here are some instructions to get a quick spectrum of your sample:
- Fill a 1mL plastic syringe approximately half way. Be sure to remove any air bubbles that may be present.
- Check to make sure a filter is properly connected and a collection flask is at the "out" port.
- Slowly inject your sample into the "in" port until you observe a few drops coming out of the "out" port.