Difference between revisions of "Walk-up NMR user instructions for Anasazi"
Jump to navigation
Jump to search
(7 intermediate revisions by the same user not shown) | |||
Line 1: | Line 1: | ||
− | If you need to use a Anasazi NMR instrument for a liquid sample, here are some | + | If you need to use a [[NMR Instrument Manufacturers|Anasazi]] NMR instrument for a liquid sample, here are some instructions to get a quick spectrum of your sample. Samples for most Anasazi instruments use regular 5mm NMR tubes. |
# Open the PNMR program. | # Open the PNMR program. | ||
# The PNMR prompt, at the bottom left-hand corner of the PNMR program, indicates the nucleus that is set to observe. It should be H1> to observe proton. If not, enter '''nu H1''' on the command line. | # The PNMR prompt, at the bottom left-hand corner of the PNMR program, indicates the nucleus that is set to observe. It should be H1> to observe proton. If not, enter '''nu H1''' on the command line. | ||
Line 13: | Line 13: | ||
# Type '''NS''' on the command line to change the number of scans, such as '''NS 16'''. | # Type '''NS''' on the command line to change the number of scans, such as '''NS 16'''. | ||
# Type '''zg''' on the command line to start the experiment. A pop-up window appears and asks for a file name, then the experiment begins. | # Type '''zg''' on the command line to start the experiment. A pop-up window appears and asks for a file name, then the experiment begins. | ||
+ | # Use the '''fo''' command to set the reference once the experiment is complete. | ||
+ | |||
+ | Here is a nice image showing a sample insertion for this instrument: | ||
+ | |||
+ | [[File:anasazi_instrument.jpg|400px]] |
Latest revision as of 03:33, 25 May 2020
If you need to use a Anasazi NMR instrument for a liquid sample, here are some instructions to get a quick spectrum of your sample. Samples for most Anasazi instruments use regular 5mm NMR tubes.
- Open the PNMR program.
- The PNMR prompt, at the bottom left-hand corner of the PNMR program, indicates the nucleus that is set to observe. It should be H1> to observe proton. If not, enter nu H1 on the command line.
- Lift the lid on the magnet to reveal the sample compartment
- Press the EJECT button on the controls and remove the old sample.
- Remove the old sample from the spinner and insert your sample.
- Insert the tube with the spinner into the black depth gauge on the front of the magnet until the button of the tube hits the bottom of the gauge.
- Place the sample/spinner on top of the probe so that it floats.
- Press INJECT on the controls and wait for the sample to drop into the magnet and start spinning.
- If the sample does not start spinning, press INJECT again, repeat until the sample spins
- Close the lid on the sample compartment
- Type shim on the command line to shim the magnet.
- Type NS on the command line to change the number of scans, such as NS 16.
- Type zg on the command line to start the experiment. A pop-up window appears and asks for a file name, then the experiment begins.
- Use the fo command to set the reference once the experiment is complete.
Here is a nice image showing a sample insertion for this instrument: