Difference between revisions of "Walk-up NMR user instructions for ThermoFisher"
Jump to navigation
Jump to search
(5 intermediate revisions by the same user not shown) | |||
Line 1: | Line 1: | ||
− | The picospin tabletop NMR instruments utilize an injection system for samples rather than the usual 5mm NMR tubes. The instrument also has a web-based interface for control and processing. | + | The [[NMR Instrument Manufacturers|picospin]] tabletop NMR instruments utilize an injection system for samples rather than the usual 5mm NMR tubes. The instrument also has a web-based interface for control and processing. Here are some instructions to get a quick spectrum of your sample: |
+ | # Fill a 1mL plastic syringe approximately half way. Be sure to remove any air bubbles that may be present. | ||
+ | # Check to make sure a filter is properly connected and a collection flask is at the "out" port. | ||
+ | # Slowly inject your sample into the "in" port until you observe a few drops coming out of the "out" port. | ||
+ | # Go to the picospin page of the web browser interface | ||
+ | # Select the '''Files''' option. | ||
+ | # Select the '''Standard Run''' option. | ||
+ | # Click '''Use these settings at run'''. | ||
+ | # The '''Run''' page will appear. | ||
+ | # Check the transmitter frequency and set the scans. | ||
+ | # Click '''Start Run'''. | ||
+ | # Process spectra with MNova, which is included with the product. | ||
+ | |||
+ | |||
[[File:picospin.jpg|400px]] | [[File:picospin.jpg|400px]] |
Latest revision as of 03:37, 25 May 2020
The picospin tabletop NMR instruments utilize an injection system for samples rather than the usual 5mm NMR tubes. The instrument also has a web-based interface for control and processing. Here are some instructions to get a quick spectrum of your sample:
- Fill a 1mL plastic syringe approximately half way. Be sure to remove any air bubbles that may be present.
- Check to make sure a filter is properly connected and a collection flask is at the "out" port.
- Slowly inject your sample into the "in" port until you observe a few drops coming out of the "out" port.
- Go to the picospin page of the web browser interface
- Select the Files option.
- Select the Standard Run option.
- Click Use these settings at run.
- The Run page will appear.
- Check the transmitter frequency and set the scans.
- Click Start Run.
- Process spectra with MNova, which is included with the product.