Difference between revisions of "Walk-up NMR user instructions for Bruker"
Jump to navigation
Jump to search
Line 3: | Line 3: | ||
==Step by step== | ==Step by step== | ||
− | # log in to workstation | + | # log in to workstation. |
− | # log in to TopSpin | + | # log in to TopSpin. |
− | # type ej on the command line or click the eject button to eject the current sample that is in the magnet | + | # type ej on the command line or click the eject button to eject the current sample that is in the magnet. |
− | # remove the old sample from the top of the magnet | + | # remove the old sample from the top of the magnet. |
− | # remove the old sample from the sample holder | + | # remove the old sample from the sample holder. |
− | # place your sample in the sample holder | + | # place your sample in the sample holder. |
− | # slide your sample down in the holder until it hits the stop | + | # slide your sample down in the holder until it hits the stop. |
− | # place the sample/holder on top of the column of air on the magnet | + | # place the sample/holder on top of the column of air on the magnet. |
− | # go back to the workstation, click insert sample and wait for the sample to drop down into the probe | + | # go back to the workstation, click insert sample and wait for the sample to drop down into the probe. |
− | # type lock on the command line | + | # type lock on the command line. |
# when the list of solvents appears, choose the one that your sample is in. Wait for locking. | # when the list of solvents appears, choose the one that your sample is in. Wait for locking. | ||
− | # once the sample is locked, type topshim on the command line to shim the sample | + | # once the sample is locked, type topshim on the command line to shim the sample. |
− | # while the sample is being automatically shimmed, type edc on the command line to create a new dataset | + | # while the sample is being automatically shimmed, type edc on the command line to create a new dataset. |
− | # edc opens the window that sets everything for the experiment to run your sample | + | # edc opens the window that sets everything for the experiment to run your sample (example [[New Dataset Window for Bruker TopSpin|here]]). |
Revision as of 10:26, 13 May 2020
Introduction
If you need to use a Bruker NMR instrument for a liquid sample, here are some quick instructions to get your sample running as easily as possible:
Step by step
- log in to workstation.
- log in to TopSpin.
- type ej on the command line or click the eject button to eject the current sample that is in the magnet.
- remove the old sample from the top of the magnet.
- remove the old sample from the sample holder.
- place your sample in the sample holder.
- slide your sample down in the holder until it hits the stop.
- place the sample/holder on top of the column of air on the magnet.
- go back to the workstation, click insert sample and wait for the sample to drop down into the probe.
- type lock on the command line.
- when the list of solvents appears, choose the one that your sample is in. Wait for locking.
- once the sample is locked, type topshim on the command line to shim the sample.
- while the sample is being automatically shimmed, type edc on the command line to create a new dataset.
- edc opens the window that sets everything for the experiment to run your sample (example here).